CARL ZEISS LSM 980 new

Location: Microscopy Room
Manufacturer: Carl Zeiss Microscopy GmbH
Model: LSM 980 with Airyscan 2
Nickname: “980”
Software: ZEN black 3.10
Year: 2024
SN:

Microscope Overview

The Carl Zeiss LSM 980 with Airyscan 2 is a cutting-edge point-scanning confocal microscope designed for generating high-resolution three-dimensional images of thick specimens. Its advanced features ensure high sensitivity, rapid imaging speeds, and minimal photodamage, making it an ideal tool for live-cell imaging and photobleaching experiments.

The inverted microscope design incorporates a dark large cage incubator and a stage-top incubator for precise temperature control and CO₂ supply, alongside Definite Focus.3 hardware to maintain focus during extended time-lapse acquisitions. A motorized stage combined with a piezo-driven Z-displacement mechanism allows for fast and precise 4D imaging across multiple positions, facilitated by the Tiles Advanced Setup within the ZEN Blue software.

Central to the LSM 980’s performance is the Airyscan 2 detector, an area detector with 32 concentrically arranged GaAsP detection elements. This innovative design allows you to acquire more of the Airy disk at once. The confocal pinhole itself remains open and does not block light, thus significantly enhancing photon collection and light efficiency. The system supports gentle super-resolution imaging with high sensitivity, achieving resolutions of up to 120 nm laterally and 350 nm axially (at 488 nm in SR mode), or as fine as 90 nm (XY) and 270 nm (Z) using Airyscan Joint Deconvolution (jDCV). The Airyscan 2 Multiplex modes (4Y and 8Y) enable parallel excitation and detection, delivering up to an 8x speed improvement without compromising sensitivity.

The LSM 980 is optimized for simultaneous spectral detection of multiple weakly labeled fluorophores, supported by lasers ranging from violet to far-red wavelengths (405, 488, 561, and 639 nm). It includes two PMT detectors and a high-sensitivity GaAsP detector, boasting a quantum efficiency of 45%, nearly double that of conventional PMTs. Additionally, the LSM Plus deconvolution mode enhances the signal-to-noise ratio by acting as a denoising step post-acquisition, enabling high-speed imaging with reduced laser power.

 

Lear more about Carl Zeiss LSM 980 here.

– MicroscopeZeiss Axio Observer

– LED IlluminationZeiss Colibri 7

– Features

                 | Definite Focus.3

                 | Airyscan 2 Multiplex

                 | LSM Plus

                 | Joint Deconvolution

System Components

Objectives

Objective

Magnification

Numerical Aperture

Medium

WD (mm)

Item Number

10x

0.3

air

5.20

420340-9901-000

20x

0.8

air

0.55

420650-9903-000

40x

1.2

Water, silicon oil or glycerine

0.41 at coverglass 0.17

420862-9970-799

63x

1.4

oil

0.19

420782-9900-799

Lasers

Laser Unit

Wavelength

Maximum Power

405

405 nm

15 mW

405

405 nm

15 mW

561

561 nm

13 mW

639

639 nm

10 mW

LEDs (Locate Mode)

Line

Wavelength

UV

385 nm

B

475 nm

G

555 nm

R

630 nm

Filter Sets (Locate Mode)

Position

Filter Set

Excitation

Beam Splitter

Emission

Reference

1

370-400 nm

450-488 nm

540-570 nm

615-648 nm

405 nm

493 nm

575 m

653 nm

410-440 nm

499-529 nm

580-605 nm

659-759 nm

FS90HE

Filter Sets (AiryScan 2))

Position

Emission Filters

Emission Dyes

1

BP 420-480 nm + BP 495-550 nm

  

2

BP 420-480 nm + BP 570-630 nm

  

3

BP 420-500 nm + LP 605 nm

  

4

BP 465-505 nm + BP 524-585 nm

  

5

BP 495-550 nm + BP 570-630 nm

  

6

BP 495-560 nm + LP 660 nm

  

7

BP 570-620 nm + LP 655 nm

  

Software

Image acquisition and analysis is performed using the Zen 3.10 software that includes the following modules:

  • z Stack
  • Time series
  • Tiles and multi-position
  • Autofocus
  • Extended focus
  • Physiology